Troubleshooting Situation #1: Meter does not enter the
test mode after inserting an AgaMatrix Presto Test Strip.
[1] CAUSE: The batteries have insufficient power. ACTION:
Replace the batteries, set time and date, and test.
[2] CAUSE: The batteries are installed incorrectly or there are
no batteries in the meter. ACTION: Check that both batteries
are installed correctly, with the positive (+) sign facing up
towards you.
[3] CAUSE: AgaMatrix Presto Test Strip has been inserted
upside down, wrong end in, or incompletely inserted into the
meter. ACTION: Insert the AgaMatrix Presto Test Strip with
the black side up and the contact bars of the test strip in the
meter. Ensure that the test strip is fully inserted.
[4] CAUSE: Defective meter or defective AgaMatrix Presto
Test Strips. ACTION: Call customer service.
[5] CAUSE: Blood or foreign objects put into the AgaMatrix
Presto Meter strip port. ACTION: Call customer service.
Troubleshooting Situation #2: The glucose test sequence
does not start after applying the blood sample.
[1] CAUSE: Defective test strip. ACTION: Repeat the test with
a new AgaMatrix Presto Test Strip. If this does not work, call
customer service.
[2] CAUSE: Sample applied after meter times out and
deactivates. ACTION: Repeat the test using a new AgaMatrix
Presto Test Strip. Wait until you see the blood and test strip
symbols on the display screen before you apply the blood
sample.
[3] CAUSE: Defective meter. ACTION: Call customer service.
Batteries
17
CHAPTER
CAUTION: When the battery symbol appears, you should
replace the batteries immediately. Only use CR2032
batteries.
Your AgaMatrix Presto Meter comes with two pre-installed,
CR2032, 3 volt, lithium batteries. If your batteries run low, the
battery symbol flashes until you change the batteries. The
backlight is disabled in all modes when the battery symbol
is on. When changing batteries the current time and date
settings will be reset. The time and date will automatically be
set to 01-01 12:00. Correct time and date settings are needed
to get correct averages. Removing the batteries does not
affect the meter’s memory or previously stored user settings,
such as those set in SET Options (e.g. alarms). If a battery is
installed incorrectly it will damage the other battery in the
meter. If this happens discard both batteries.
Note: Dispose of used batteries according to your local environ-
mental regulations. Your meter is an electronic device. When dis-
posing of your meter follow all local environmental regulations.
Replacing the Batteries
[1] Open Battery Door: Make
sure the meter is deactivated. Turn
the meter so you are looking at
the battery door. Press the battery
door clip to open. Remove the